UV-Photo-X Appendix

Teledyne

Analytical

Instruments

31

The indicator LEDs are located to the left of the display
along the outer edge of the Fluorometer housing. Refer to
Figure 3-1. If the indicators are non-green, refer to the
Diagnostics section in Section 5. Preconfiguration at the
TIA factory should eliminate the occurrence of non-green
indicators.

6) Perform Mode 2 initialization

a) Press the INT PRESET, located in the upper left

hand corned of the Fluorometer (refer to Figure 3-1),
for at least five (8) seconds.

b) Both the REF and MSR DET LEDs exhibit solid

green or flashing green output.

c) The value on the display is the fluorescence level

due to the probe/analyzer configuration. Record this
value for later analysis. Table 1 below provides an
example of how to organize measurement data.

d)

Insert the span filter using the SPN CHK button
located on the upper end of the Fluorometer
housing. When the filter is in place, the FLTR-IN
LED will change from green to red.

e) Wait 2 minutes for the output to stabilize and the

unit should read an upscale value.

f) Both the REF and MSR DET LEDs should have

solid green or flashing green output.

g) If the indicator LED color does change color, refer

to the Diagnostics section below.

h) Depress the SPN CHK button to remove the filter

from the beam path. The FLTR-IN indicator will
change from red to green.

i) Wait 2 minutes for the output to stabilize and the

unit should display residual fluorescence level of the
probe/analyzer configuration.

j) Both the REF and MSR DET LEDs should have

solid green or flashing green output.

k) The unit is now ready for either application

engineering studies or process monitoring.