UV-Photo-X Specifications

Teledyne

Analytical

Instruments

27

6.2 Transmitter

Specifications

1) Configuration.XSource Reference Through Media

Reference.

Application specific settings
Operation mode Display format XXXX

specified 4mA = (unspec, field settable)
specified 20mA = (unspec, field settable)

2) Optical specifications.

a) Lamp.

Xenon Flash Lamp, Life: 3-5 years

b) Detectors.

PMT

Spectral range: UV 190-690 nm

3) Electrical specifications. ( Tungsten Halogen, X Flash

Lamp)

a) Power Requirements 24VDC(IEC connector)

Current Requirements; LED ~332mA Tungsten
Halogen ~728.5mAmA (24VDC)

Consumption:

LED 8 W Tungsten Halogen ~17.5W (24VDC)

b) Output signal: Isolated 4-20mA sourced output

Permissible loop resistance
500 ohms for customer use

c) Span calibration:

Remote trigger via normally closed contact (must remove the
jumper on the top of the Photo-X unit for remote triggering)