Specifications Model

6650

26 Teledyne

Analytical

Instruments

Specifications

6.1 General Measurement Specifications

1) Measurement method.

OPTICAL METHOD: Fluorescence

Analysis type: Front Surface Fluorescence

2) Sampling Wavelengths.

Excitation: UG11

Reference: 328nm

bandpass

Emission:

420 nm UVLP

3) Optical path length.

NA.

4) Sample.

Mode 1 Calibration

Salcon Water = 455 counts

5ppm in water = 1000 counts

Water + span filter = OFF-SCALE

5ppm + span filter = OFF-SCALE

Mode 2 Calibration

Salcon Water = 044 counts

5ppm in water = 081counts

Span filter = 279 counts

5) Optical span filter.

Span Calibration value: see above

Span Filter Type: SS-KFLY-5