beautypg.com

Definition of terms – Teledyne 6650 - Fluorescence Probe - Oil in Water analysis system User Manual

Page 12

background image

Definition of Terms

Model 6650

4 Teledyne

Analytical

Instruments

Definition of Terms

2.1 Excitation

Filter

The excitation filter is used to select the range of wavelengths, the

pass band, used to cause or excite the molecular fluorescence.
Wavelengths not in the pass band are rejected and ideally never reach
the sample. Removal of wavelengths outside the pass band minimizes
the possibly of false fluorescence readings due to detection of light from
the lamp which mimics the fluorescence signal.

2.2 Emission

Filter

The emission filter is used to select the range of wavelengths, the

pass band, to be passed to the measure detector. Wavelengths not in the
pass band are rejected and ideally never reach the sample. It is critical
that the excitation wavelengths never reach the detector, since it will
respond to excitation light. Excitation light impinging the measure
detector results in increased residual background levels, which reduces
the dynamic range, the signal-to-noise ratio and the signal-to-blank ratio.

2.3 Measure

Detector

The light detector is most often a photomultiplier tube, though

photodiodes are increasingly being used. The light passing through the
emission filter is detected by the photomultiplier or photodiode. The
light intensity, which is proportional to the analyte concentration, is
registered as a digital readout.

2.4 Span Filter

The span filter is used to check instrument operation. When a

fluorescence filter is employed the span filter fluorescence value is
dependent on the initialization method utilized and the standards used
during calibration. The span filter is used to verify instrument operation
and should not be used to calibrate the instrument.