Operation Series

4040

60 Teledyne

Analytical

Instruments

Pressing enters the SPAN mode in the NMH mode first.

The first pass in the NMH mode only purges the sample gas from the
column (no span calculation is performed this time around). Then the
CH4 is span calibrated followed by the second pass of the NMH mode,
where the span calculation is performed.

4.13 The Alarms Function

The Model 4040 is equipped with two alarms and a system failure

alarm relay. Each alarm relay has a set of form "C" contacts rated for 3
amperes resistive load at 250 VAC. See Figure in Chapter 3, Installation
and/or the Interconnection Diagram included at the back of this manual
for relay terminal connections. The alarm relay contacts are accessible to
the user from the 50-pin Equipment Interface connector.

The system failure alarm has a fixed configuration described in

Chapter 3 Installation.

The concentration alarms can be configured from the ALARM

function screen as follows:
________________________________________________

-> AL-1 DEFEATED

AL-1 HIGH

AL-1 NON-FAILSAFE

AL-1 NON-LATCHING

AL-1 20.00 ppb

AL-1 SOURCE CH4

AL-2 ACTIVE

AL-2 LOW

AL-2 FAILSAFE

AL-2 LATCHING

AL-2 50.00 ppb

AL-2 SOURCE CH4

________________________________________________

Note: For the purpose of example, Alarm 2 has been set differently

than Alarm 1.)

Note: An alarm in 'alarm condition' is signaling that action must be

taken such as correcting the alarm or the analysis