Operation Series

4040

48 Teledyne

Analytical

Instruments

- Range
- Activity

Indicator

The MAIN MENU and SUB MENU screens have a flashing arrow on
the left side indicating which item is being pointed to.

MODIFIABLE ITEM

screens have a flashing value somewhere in the

line.

4.5.1 Main Menus

The MAIN MENU consists of 13 functions you can use to customize

and check the operation of the analyzer. They are listed here with brief
descriptions:

1.

AUTO-CAL:

Used to define and/or start an automatic calibration

sequence. AUTOCAL is not available for standard
configurations of the 4040.

2.

GAIN:

Used to change the amplification of the incoming signal.

This is factory set, and should not change.

3.

PASSWORD:

Used to establish password protection or change

the existing password.

4.

LOGOUT:

Logging out prevents unauthorized tampering with the

analyzer settings.

5.

MODEL:

Displays Manufacturer, Model, and Software version of

the instrument.

6.

SELF-TEST:

The instrument performs a self-diagnostic routine to

check the integrity of the power supply, output boards, cell and
amplifiers.

7.

SPAN:

Set up and/or start a span calibration

8.

ALARMS:

Used to set the alarm setpoints and determine whether

each alarm will be active or defeated, HI or LO acting, and
failsafe or not.

9.

RANGE:

Used to set up three analysis ranges that can be

switched automatically. with auto-ranging or used as individual
fixed ranges.

10.

ANALOG-OUT ADJUST: Adjust the analog output with offset and
scaling for Channel 1 only.

11. CHANGE STREAM: Controls relay switch for alternate gasses.

CHANGE STREAM may not be set during a span calibration.