6 setting up an auto-cal, 7 adjusting the gain – Teledyne 4040 - Methane / Nonmethane analyzer User Manual

Page 53

background image

Operation

Teledyne

Analytical

Instruments

53

-----------------------

STANDBY

Switch to Sleep Mode

ENTER-YES ESCAPE-NO

notification if logged out

-----------------------

Access

Restricted

Re-enter

Password

-----------------------

4.6 Setting up an AUTO-CAL

AUTOCAL is not available for standard configurations of the

4040.

4.7 Adjusting the GAIN


The output of the FID is conditioned by an amplification circuit with

3 user-adjustable settings:

0) x1
1) x10
2) x100

The resulting signal is converted to a number. Numbers outside the

range –262144 to 262143 are unreliable and lead to inaccurate
measurements. If the number is outside this range during analysis, the
message 'GAIN TOO HIGH' is shown on the 2

nd

line of the display, and

the analyzer is not updated to a new, calculated value. The Gain Too
High message may be removed by pressing the ESCAPE key.

The gain has been set at the factory such that the analyzer will work

reliably within the prescribed range of operation. The user may change
the gain when a different level of sample is to be analyzed.