beautypg.com

Teledyne 4040 - Methane / Nonmethane analyzer User Manual

Page 35

background image

Installation

Teledyne

Analytical

Instruments

35

3.3.4 RS-232 Port

The digital signal output is a standard RS-232 serial communications

port used to connect the analyzer to a computer, terminal, or other digital
device. It requires a standard 9-pin D connector.

Input: The input functions using RS-232 that have been implemented
to date are described in Table 3-7.

Table 3-8: Commands via RS-232 Input

Command Description

AS

Immediately starts an autospan.

AL

Immediately revert to Analyze (CH4 Mode)

RL

Change to Manual Range LOW

RM

Change to Manual Range MEDIUM

RH

Change to Manual Range HIGH

RA

Change to Manual Range AUTO

Implementation: The RS-232 protocol allows some flexibility in its
implementation. Table 3-9 lists certain RS-232 values that are required
by the Model 4020 implementation.

Table 3-9: Required RS-232 Options

Parameter Setting

Baud

2400

Byte

8

bits

Parity

none

Stop

Bits

1

Message Interval

Sent at the end of each cycle.

3.3.5 Supporting Gases

Normally, four supporting gases of different composition (see

Section 4.1: Equipment) will be required to operate the analyzer. The
recommended composition of these gases is specified in the Application
Data section of the Appendix. The gases should be supplied from
cylinders that are equipped with the type of regulator specified in the
aforementioned sections.