Installation

Teledyne

Analytical

Instruments

31

Threshold Alarm 2:
Can be configured as high (actuates when concentration is above
threshold), or low (actuates when concentration is below threshold).
Configured as failsafe or non-failsafe
Configured as non-latching
Can be configured out (defeated).
Can be set to trigger by levels of CH4, NMH, or THC.

System Alarm:

Actuates when power is removed from analyzer.
It is triggered if 4040 fails to ignite after five times.

Note: (Reset by pressing the STANDBY button to remove power.

Then press STANDBY again and any other button except
SYSTEM to resume. Further detail can be found in Chapter 4,
Section 4.5.15.

Table 3-4: Alarm Relay Contact Pins

Pin Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Threshold Alarm 2, normally closed contact

44

Threshold Alarm 2, moving contact

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact