3.3.3.7

R

ANGE

ID

R

ELAYS

There are three dedicated Range ID relay contacts. They are

assigned to relays in ascending order—Low range is assigned to Range
1 ID, Medium range is assigned to Range 2 ID, and High range is
assigned to Range 3 ID. Table 3-6 lists the pin connections. Contacts
are normally open, and they close when 4040 switches to that particular
range.

Table 3-6: Range ID Relay Connections

Pin

Function

21

Range 1 ID Contact

38

Range 1 ID Contact

22

Range 2 ID Contact

39

Range 2 ID Contact

19

Range 3 ID Contact

18

Range 3 ID Contact

34

Not

Used

35

Not

Used

3.3.3.8

N

ETWORK

I/O

A serial digital input/output for local network protocol. At this

printing, this port is not yet functional. It is to be used in future options
to the instrument. Pins 13 (+) and 29 (–).

3.3.3.9

P

IN

O

UT

T

ABLE

The following table summarizes all the outputs/inputs available in

the 50 pin D-Sub connector on the back panel of the Analyzer.

Table 3-7: Pin out of 50 pin D-Sub Connector

pin #

Description

1

2

3 + Output 4-20 ma - Channel 2
4 - Output 4-20 ma - Channel 2
5 + Output 4-20 ma – Channel 1
6 - Output 4-20 ma – Channel 1
7 - Output 0-1 v (Channel 1)
8 + Output 0-1 v (Channel 2)
9