beautypg.com

Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 79

background image

Oxygen Analyzer

Part II: Analysis Unit

Part II: 4-3

Teledyne Analytical Instruments

Table 4-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Preamp

0

OK

1

Zero too high

2

Amplifier output doesn't match test input

3

Both Failed

The results screen alternates for a time with:

Press Any Key
To Continue...

Then the analyzer returns to the initial System screen.

4.5

Cell Failure Checks

Cell failure is covered in detail in Part I: Control Units, section 4.4.1.3,

Cell Failure. Cell replacement is covered Part II: Analysis Units chapter 5,
Maintenance.

Cell failure in the 3010T is usually associated with inability to zero the

instrument down to a satisfactorily below 5PPM reading. When this occurs,
the 3010 system alarm trips, and the VFD displays a failure message.

#.# ppm Anlz

CELL FAIL/ ZERO HIGH

Before replacing the cell:

a. Check your span gas to make sure it is within specifications.

b. Check for leaks downstream from the cell, where oxygen may be

leaking into the system.