Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 61

background image

Part II: 1-3

Oxygen Analyzer

Part II: Analysis Unit

Teledyne Analytical Instruments

Figure 1-2: Electrical Connector/Control Panel

Power In

Power input terminals for electric heater.
Requires 115 or 230 V ac, depending on
position of the Voltage Selector switch. Use
50/60 Hz.

CAUTION: Check the position of the Voltage Selector switch

BEFORE applying power to the Power Input termi-
nals.

Voltage Selector Power input selector switch for electric

heater. Adjusts input requirement for 115 or
230 V ac, depending on available source
voltage. Use 50/60 Hz.

Fuses

1.6 A, 250 V, T type, European size
5

× 20 mm fuses. Fuse 1 is on the neutral

side of the line. Fuse 2 is on the hot side of
the line.

Solenoid Valves Terminals that provide all electrical intercon-

nections from the Control Unit to the gas
control valves.

Sensor Signal

Terminals that provide connections from the
Micro-Fuel Cell sensor to the Control Unit.

115V