beautypg.com

Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 70

background image

3 Installation

Model 3010T

3-2: Part II

Teledyne Analytical Instruments

Figure 3-1 is a view with the cover removed showing the external Gas

Connector Panel and the internal Electrical Connector Panel.

Figure 3-1: View of Analysis Unit Showing Connector Panels

3.3

Gas Connector Panel Connections

Before using this instrument, it should be determined if the unit will be

used for pressurized service or vacuum service and low pressure applica-
tions. Inspect the restrictor kit that came with the unit. The kit consist of two
restrictors and a union for 1/4” diameter tubing. Notice that the two 1 3/4”
long, 1/4” diameter tubing are restrictors. It has an open end and a closed
end with a small circular orifice. The restrictor without the blue sticker is for
;ow pressure and vacuum service. For high pressure (5 to 50 psig) applica-
tions, use the restrictor that has a blue sticker on the body.