beautypg.com

Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 38

background image

4 Operation

Model 3010TB

4-10: Part I

Teledyne Analytical Instruments

Press the

System

button to start the

System

function.

TRAK/HLD Auto—Cal
PSWD Logout More

Use the < > arrow keys to blink More, then press

Enter

.

Version Self—Test

Use the < > arrow keys again to move the blinking to the Self–Test

function. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC
Testing Preamp — 83

During preamp testing there is a countdown in the lower right corner of

the screen. When the testing is complete, the results are displayed.

Power: OK Analog: OK
Preamp: 3

The module is functioning properly if it is followed by OK. A number

indicates a problem in a specific area of the instrument. Refer to chapter 5
Maintenance
for number-code information. The results screen alternates for a
time with:

Press Any Key
To Continue...

Then the analyzer returns to the initial System screen.

4.3.6 Version Screen

Move the < > arrow key to More and press

Enter

. With Version

blinking, press

Enter

. The screen displays the manufacturer, model, and

software version information.

4.3.7 Showing Negative Oxygen Readings

For software version 1.4.4 or later, the instrument only displays oxygen

readings that are positive or zero. The instrument can be reconfigured to
show negative readings if sensor output drifts below zero. This situation
may arise after the instrument has been zeroed, as time progresses the sensor
may drift below the zero calibration setpoint.

To show negative oxygen readings on the display:

- Press the System key