beautypg.com

Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 25

background image

Oxygen Analyzer

Part I: Control Unit

Part I: 3-9

Teledyne Analytical Instruments

4) When the CRC closes, remove the span command.

When CRC opens again, zero and span are done, and the sample is

being analyzed.

Note: The Remote Probe connector (paragraph 3.3) provides signals

to ensure that the zero and span gas valves will be controlled
synchronously. If you have the -C Internal valve option -
which includes additional zero and span gas inputs - the
3010TB automatically regulates the zero, span and sample gas
flow.

Range ID Relays: Four dedicated RANGE ID CONTACT relays .

The first three ranges are assigned to relays in ascending order—Low range
is assigned to RANGE 1 ID, Medium range is assigned to RANGE 2 ID,
and High range is assigned to RANGE 3 ID. RANGE 4 ID is reserved for
the Air Cal Range (25%).

Network I/O: A serial digital input/output for local network protocol.

At this printing, this port is not yet functional. It is to be used in future
versions of the instrument.

RS-232 Port: The digital signal output is a standard RS-232 serial

communications port used to connect the analyzer to a computer, terminal, or
other digital device. The pinouts are listed in Table 3-3.

Table 3-3: RS-232 Signals

RS-232 Sig RS-232 Pin Purpose

DCD

1

Data Carrier Detect

RD

2

Received Data

TD

3

Transmitted Data

DTR

4

Data Terminal Ready

COM

5

Common

DSR

6

Data Set Ready

RTS

7

Request to Send

CTS

8

Clear to Send

RI

9

Ring Indicator

The data sent is status information, in digital form, updated every two

seconds. Status is reported in the following order: