beautypg.com

Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 76

background image

3 Installation

Model 3010T

3-8: Part II

Teledyne Analytical Instruments

Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.

Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors

Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.

Power up the system, and test it as follows:

1. Repeat the Self-Diagnostic Test as described in Part I, chapter 4,

section 4.3.5.