beautypg.com

Teledyne 3010MB - Split architecture paramagnetic oxygen analyzer User Manual

Page 69

background image

A-5

Percent Paramagnetic Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

A-5

Zero Cal

The

ZERO

function on the MAIN MENU is used to enter the zero

calibration function. Zero calibration can be performed in either the auto-
matic or manual mode. In the automatic mode, an internal algorithm com-
pares consecutive readings from the sensor to determine when the output is
within the acceptable range for zero. In the manual mode, the operator
determines when the reading is within the acceptable range for zero.

Make sure the zero gas is connected to the instrument.

If you have trouble zeroing, you may need to skip to section 4.8.1.3

Cell Failure.

Auto Mode Zeroing

Select

ZERO

to enter the

ZERO

function. The

ZERO

screen allows

you to select whether the zero calibration is to be performed automatically or
manually. Use the DOWN/UP control to toggle between

AUTO

and

MAN

zero settling. Stop when

AUTO

appears on the display.

Zero: Settling: AUTO

To Begin

ENTER to begin zeroing.

#### % Zero

Slope=#### ppm/s

The beginning zero level is shown in the upper left corner of the dis-

play. As the zero reading settles, the screen displays and updates information
on

Slope

(unless the Slope starts within the acceptable zero range and does

not need to settle further).

Then, and whenever Slope is less than 0.08 for at least 3 minutes,

instead of Slope you will see a countdown, such as

1 Left

,

0 Left

. These are

steps in the zeroing process that the system must complete, AFTER settling,
before it can go back to ANALYZE.

#### % Zero

1 Left=### ppm/s

The zeroing process will automatically conclude when the output is

within the acceptable range for a good zero. Then the analyzer automatically
returns to the ANALYZE screen.