beautypg.com

Range – Teledyne 3010MB - Split architecture paramagnetic oxygen analyzer User Manual

Page 48

background image

4 Operation

Model 3010MB

4-18: Part I

Teledyne Analytical Instruments

AL—1 AL—2

Choose Alarm

Set up alarm 1 by moving the blinking over to

AL–1 using the < >

arrow keys. Then press

Enter

to move to the next screen.

AL—1 1000 ppm HI

Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

Value of the alarm setpoint, AL–1 #### (ppm or % oxygen)

Out-of-range direction, HI or LO

Defeated? Dft–Y/N (Yes/No)

Failsafe? Fs–Y/N (Yes/No)

Latching? Ltch–Y/N (Yes/No).

To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the

∆∇

arrow keys to

change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, setpoint units are
parts-per-million.)

To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the

∆∇

arrow

keys to change the parameter.

Once the parameters for alarm 1 have been set, press

Alarms

again, and repeat this procedure for alarm 2 (AL–2).

To reset a latched alarm, go to

Dft– and then press either

two

times or

two times. (Toggle it to

YYYYY and then back to N.

N.

N.

N.

N.)

–OR –

Go to

Ltch– and then press either

two times or

two times.

(Toggle it to

N

NN

N

N and back to Y.

Y.

Y.

Y.

Y.)

4.6

The

Range

Function

The Range function allows the operator to program up to three concen-

tration ranges to correlate with the DC analog outputs. If no ranges are
defined by the user, the instrument defaults to:

Range

Limits

Low

0–1%

Med

0–5%

High

0–10%