Percent Paramagnetic Oxygen Analyzer

Part I: Control Unit

Part I: 4-9

Teledyne Analytical Instruments

If no alarms are tripped, the

Analyze

screen appears as:

0.0 ppm AnlZ

Range: 0 — 100

If an alarm is tripped, the second line will change to show which alarm

it is:

0.0 ppm Anlz

A L — 1

NOTE:If you previously logged off the system , you will now be

required to re-enter the password to gain access to Span,
Zero, Alarm, and Range functions.

4.3.4 Logout

The

Logout function provides a convenient means of leaving the

analyzer in a password protected mode without having to shut the instrument
off. By entering

Logout, you effectively log off the instrument leaving the

system protected against use until the password is reentered. To log out,
press the

System

button to enter the

System

function.

TRAK/HLD Auto—Cal

PSWD Logout More

Use the < > arrow keys to position the blinking over the

Logout func-

tion, and press

Enter

to Log out. The screen will display the message:

Protected Until

Password Reentered

4.3.5 System Self-Diagnostic Test

The Model 3010MB has a built-in self-diagnostic testing routine. Pre-

programmed signals are sent through the power supply, output board and
sensor circuit. The return signal is analyzed, and at the end of the test the
status of each function is displayed on the screen, either as

OK or as a

number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for
number code.)

Note: Remote Probe connector must be connected to the Analysis

Unit, or sensor circuit will not be properly checked.

The self diagnostics are run automatically by the analyzer whenever the

instrument is turned on, but the test can also be run by the operator at will.
To initiate a self diagnostic test during operation: