Percent Paramagnetic Oxygen Analyzer
Part I: Control Unit
Part I: 4-19
Teledyne Analytical Instruments
The Model 3010MB is set at the factory to default to autoranging. In
this mode, the microprocessor automatically responds to concentration
changes by switching ranges for optimum readout sensitivity. If the current
range limits are exceeded, the instrument will automatically shift to the next
higher range. If the concentration falls to below 85% of full scale of the next
lower range, the instrument will switch to that range. A corresponding shift
in the DC percent-of-range output, and in the range ID outputs, will be
noticed.
The autoranging feature can be overridden so that analog output stays
on a fixed range regardless of the oxygen concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 V dc (20 mA at the current output).
However, the digital readout and the RS-232 output of the concentra-
tion are unaffected by the fixed range. They continue to read accurately with
full precision. See Front Panel description in chapter 1.
The automatic air calibration range is always 0-25 % and is not pro-
grammable.
4.6.1 Setting the Analog Output Ranges
To set the ranges, enter the range function mode by pressing the
Range
button on the front panel.
L### M####
H##### ModeAUTO
Use the < > arrow keys to blink the range to be set: low (
L), medium
(
M), or high (H).
Use the
∆∇
arrow keys to enter the upper value of the range (all ranges
begin at 0 ppm). Repeat for each range you want to set. Press
Enter
to
accept the values and return to
Analyze
mode. (See note below.)
Note: The ranges must be increasing from low to high, for example,
if range 1 is set as 0–10% and range 2 is set as 0–100%, range
3 cannot be set as 0–50% since it is lower than range 2.