Percent Paramagnetic Oxygen Analyzer

Part I: Control Unit

Part I: 4-19

Teledyne Analytical Instruments

The Model 3010MB is set at the factory to default to autoranging. In

this mode, the microprocessor automatically responds to concentration
changes by switching ranges for optimum readout sensitivity. If the current
range limits are exceeded, the instrument will automatically shift to the next
higher range. If the concentration falls to below 85% of full scale of the next
lower range, the instrument will switch to that range. A corresponding shift
in the DC percent-of-range output, and in the range ID outputs, will be
noticed.

The autoranging feature can be overridden so that analog output stays

on a fixed range regardless of the oxygen concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 V dc (20 mA at the current output).

However, the digital readout and the RS-232 output of the concentra-

tion are unaffected by the fixed range. They continue to read accurately with
full precision. See Front Panel description in chapter 1.

The automatic air calibration range is always 0-25 % and is not pro-

grammable.

4.6.1 Setting the Analog Output Ranges

To set the ranges, enter the range function mode by pressing the

Range

button on the front panel.

L—### M—####

H—##### Mode—AUTO

Use the < > arrow keys to blink the range to be set: low (

L), medium

(

M), or high (H).

Use the

∆∇

arrow keys to enter the upper value of the range (all ranges

begin at 0 ppm). Repeat for each range you want to set. Press

Enter

to

accept the values and return to

Analyze

mode. (See note below.)

Note: The ranges must be increasing from low to high, for example,

if range 1 is set as 0–10% and range 2 is set as 0–100%, range
3 cannot be set as 0–50% since it is lower than range 2.