beautypg.com

Teledyne 3010MB - Split architecture paramagnetic oxygen analyzer User Manual

Page 54

background image

5 Maintenance

Model 3010MB Percent Paramagnetic Oxygen Analyzer

5-2: Part I

Teledyne Analytical Instruments

Figure 5-1: Removing Fuse Block Cap and Fuse from Housing

2. Replace fuse by reversing process in step 1.

5.2

System Self Diagnostic Test

1. Press the

System

button to enter the system mode.

2. Use the < > arrow keys to move to More, and press

Enter

.

3. Use the < > arrow keys to move to Self-Test, and press

Enter

.

The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failures

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed