4: part ii, 40 % of n, And 50% of ar. c0 – Teledyne 3010MB - Split architecture paramagnetic oxygen analyzer User Manual
Page 62
1-4: Part II
1 Operational Theory
Model 3010M
Teledyne Analytical Instruments
Gas
Cross
Gas
Cross
Sensitivity
Sensitivity
in vol. %
in vol. %
Acetylene C
2
H
2
-0.24
Hydrogen chloride HCI
-0.30
Allene C
3
H
4
-0.44
Hydrogen fluoride HF
+0.10
Ammonia NH
3
-0.26
Hydrogen iodide HI
-1.10
Argon Ar
-0.22
Hydrogen sulphide H
2
S -0.39
Bromine Br
2
-1.30
Iodine I
-2.40
1.2 Butadiene C
4
H
6
-0.49
Isobutane C
4
H
10
-1.11
1.3 Butadiene C
4
H
6
-0.49
Isopantane C
5
H
12
-1.49
n-Butane C
4
H
10
-1.11
Krypton Kr
-0.51
I-Butane C4H8
-085
Laughing gas N
2
O
-0.20
cis 2-Butane C
4
H
8
-0.89
Methane CH
4
-0.20
trans 2-Butane C
4
H
8
-0.92
Neon Ne
+0.13
Carbon dioxide CO
2
-0.27
Neopentane C
5
H
12
-1.49
Carbon monoxide CO
+0.06
Nitric acid HNO
3
+0.43
Chlorine Cl
2
-0.77
Nitrogen dioxide NO
2
+28.00
Cyclo hexane C
6
H
12
-1.56
Nitrous oxide NO
+40.00
Ethane C
2
H
6
-0.43
n-Octane C
8
H
18
-2.50
Ethylene C
2
H
4
-0.26
n-Pentane C
5
H
12
-1.45
Helium He
+0.30
Propane C
3
H
8
-0.86
n-Heptane C
7
H
16
-2.10
Propylene C
3
H
6
-0.55
n-Hexane C
6
H
14
-1.70
Vinyl chloride
-0.63
Hydrogen H
2
+0.24
Water H
2
O
-0.02
Hydrogen bromide HBr
-.0.61
Xenon Xe
-0.95
With gas mixtures the components are weighted according to their
proporational volumes.
The cross interference with a gas mixture can be determined in advance if the
concentration of the individual background gases is known.
Example:
Cross sensitivity calculation for a gas mixture:
The gas to be measured consists of 10 % CO
2
, 40 % of N
2
, and 50%
of Ar.
C0
2
: -0.27 x 10% = -0.027
Ar: -0.22x5O%= -0.11
TOTAL CROSS INTERFERENCE -0.137
Calculation of the zero value:
(0% 0
2
) - (Cross Interference Value) = Zero Point Value
0% -(-0.137) = +0.137%0
2