3 Installation Model 3010MB

3-4: Part I

Teledyne Analytical Instruments

Primary Input Power:

The universal power supply requires a 100-

240V ac, 50/60 Hz power source. See Figure 3-4 for detailed connections.

DANGER: Power is applied to the instrument's circuitry as

long as the instrument is connected to the power
source. The standby function switches power on or
off to the displays and outputs only.

Figure 3-4: Primary Input Power Connections

Fuse Installation:

The fuse holders accept 5 x 20 mm, 1.0 A, T

type (slow blow) fuses. Fuses are not installed at the factory. Be sure to
install the proper fuse as part of installation (See Fuse Replacement in
chapter 5, maintenance.)

Analog Outputs:

There are eight DC output signal connectors on

the ANALOG OUTPUTS terminal block. There are two connectors per
output with the polarity noted. See Figure 3-5.

The outputs are:

0–1 V dc % of Range: Voltage rises linearly with increasing oxygen, from

0 V at 0% to 1 V at 100%. (Full scale = 100%
programmed range.)

0–1 V dc Range ID:

0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.