Maintenance 3000ZA-3X


Teledyne

Analytical

Instruments

48

2. To change between American and European fuses, remove

the single retaining screw then flip the Fuse Block over 180
degrees. Replace the screw.

3. Replace fuse as shown in Figure 5-3.
4. Reassemble housing as shown in Figure 5-2.

American Fuses

European Fuses

Figure 5-3: Installing Fuses

5.4 System Self Diagnostic Test

1. Press the SYSTEM button to enter the system mode.
2. Use the ◄► arrow keys to move to More, and press

ENTER.

3. Use the ◄► arrow keys to move to Self-Test, and press

ENTER.

The following Items are tested:

Table 5-1: Self-Test Result Display

“5V - Test : GOOD”

( +5VDC power supply )

“15V - Test : GOOD”

(+/-15 VDC power supply )

“DAC A - Test : GOOD”

(% of range analog output )

“DAC B - Test : GOOD”

( Range ID analog output )

“Pre L - Test : GOOD” ( Tests offset of low range amplifier )

“Pre M - Test : GOOD” ( Tests offset of medium range amplifier)

“Pre H - Test : GOOD” ( Tests offset of high range amplifier )