Operation 3000ZA-3X


Teledyne

Analytical

Instruments

42

Pressing ENTER causes the word HIGH or LOW to start flashing.

Then pressing UP or DOWN toggles between the 2 words. When HIGH,
the alarm will trip if O

2

concentration is above the setpoint as discussed

below.

4.6.3 Alarm Failsafe

-----------------------

->AL-1

FAILSAFE

-----------------------
Pressing ENTER causes the word FAILSAFE or NON-FAILSAFE to

start flashing. Then pressing UP or DOWN toggles between the 2
words. When FAILSAFE, the alarm relay will be energized in a non-
tripped state, and de-energized when tripped.

4.6.4 Alarm Latching

-----------------------

->AL-1

LATCHING

-----------------------
Pressing ENTER causes the word LATCHING or NON-LATCHING to

start flashing. Then pressing UP or DOWN toggles between the 2
words. When LATCHING, an alarm will remain tripped until defeated;
even if the O

2

concentration returns to a level that does not trip the

alarm. When NON-LATCHING, a tripped alarm will become un-tripped if
the O

2

concentration returns to a non-tripping level.

4.6.5 Alarm Setpoint

-----------------------

->AL-1

80.00

ppm

-----------------------
Pressing ENTER causes the setpoint concentration to start flashing.

Then pressing UP or DOWN adjusts the setpoint up or down
respectively. Setpoints may range from 0.00 ppm to 100.00 %.