Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

19

4–20 mA DC % Range: Current increases linearly with

increasing oxygen, from 4 mA at 0
ppm to 20 mA at full scale ppm.
(Full scale = 100% of
programmable range)

4–20 mA DC Range ID: 7.2 mA = Low Range

12 mA = Medium Range
16.8 mA = High Range

Table 3-1: Analog Output Connections

Pin

Function

3

+ Range ID, 4-20 mA, floating

4

– Range ID, 4-20 mA, floating

5

+ ppm Range, 4-20 mA, floating

6

– ppm Range, 4-20 mA, floating

8

+ Range ID, 0-1 VDC

23

– Range ID, 0-1 V dc, negative ground

24

+ ppm Range, 0-1 VDC

7

– ppm Range, 0-1 VDC, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 VAC into a
resistive load. The connectors are:

Threshold Alarm 1:

Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).

Can be configured as failsafe or non-failsafe.

Can be configured as latching or non-latching.

Can be configured out (defeated).