beautypg.com

Teledyne 3000ZA-3X - Trace oxygen analyzer User Manual

Page 21

background image

Trace Oxygen Analyzer

Operational Theory

Teledyne

Analytical

Instruments

9

Figure 2-2: Flow Diagram with Options

2.4 Zirconium Oxide Sensor Application Notes

The following sample gases require conditioning prior to entering

the sensors:

Flammable gases such as methane, alcohol and carbon
monoxide may cause a measurement error. They must be
filtered from the sample gas before entering the cell.

Gases containing halogen atoms (F, Cl, Br, etc.) such as
Freon, must be filtered before entering the cell. The sensor
will be damaged by the decomposition of Freon.

Gases containing SOx, H2S, silicone vapor and adhesives
must not enter the sensor. These components adversely affect
the performance of the sensor.

Dust and oil should be filtered from the sample gas.

Water vapor in contact with the sensor will damage the
sensor. Use a moisture trap or filter to remove any water
vapor in the sample.

2.5 Electronics and Signal Processing

The Model 3000ZA Trace Oxygen Analyzer uses an 8031

microcontroller with 32 kB of RAM and 128 kB of ROM to control all
signal processing, input/output, and display functions for the analyzer.