Threshold Alarm 2:
Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
System
Alarm:
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be defeated.
Actuates if self test fails.
(Reset by pressing button to remove power to relay.
Then press again and any other button Up and Down
simultaneously to resume.
Further detail can be found in Chapter 4, Section 4.6.
Table 3-2: Alarm Relay Contact Pins
Pin Contact
45
Threshold Alarm 1, normally closed contact
28
Threshold Alarm 1, moving contact
46
Threshold Alarm 1, normally open contact
42
Threshold Alarm 2, normally closed contact
44
Threshold Alarm 2, moving contact
43
Threshold Alarm 2, normally open contact
36
System Alarm, normally closed contact
20
System Alarm, moving contact
37
System Alarm, normally open contact
Digital Remote Cal Inputs: Accept 0 V (off) or 24
VDC (on) inputs for
remote control of calibration. (See Remote Calibration Protocol below.)
See Table 3-3 for pin connections.