Teledyne 3000ZA-3X - Trace oxygen analyzer User Manual

Page 32

background image

Installation 3000ZA-3X


Teledyne

Analytical

Instruments

20

Threshold Alarm 2:

Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).

Can be configured as failsafe or non-failsafe.

Can be configured as latching or non-latching.

Can be configured out (defeated).

System

Alarm:

Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be defeated.
Actuates if self test fails.

(Reset by pressing button to remove power to relay.
Then press again and any other button Up and Down
simultaneously to resume.

Further detail can be found in Chapter 4, Section 4.6.

Table 3-2: Alarm Relay Contact Pins

Pin Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Threshold Alarm 2, normally closed contact

44

Threshold Alarm 2, moving contact

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact

Digital Remote Cal Inputs: Accept 0 V (off) or 24

VDC (on) inputs for

remote control of calibration. (See Remote Calibration Protocol below.)
See Table 3-3 for pin connections.