Installation 3000ZA-3X


Teledyne

Analytical

Instruments

24

Figure 3-6: FET Series Resistance

3.3.2.3

RS-232

P

ORT

The digital signal output is a standard, full duplex RS-232 serial

communications port used to connect the analyzer to a computer,
terminal, or other digital device. It requires a standard 9-pin D
connector.

The output data is status information, in digital form, updated every

two seconds. Status is reported in the following order:

The concentration in ppm or percent

The range in use (HI, MED, LO)

The span of the range (0-100 ppm, etc)

Which alarms—if any—are disabled (AL–x DISABLED)

Which alarms—if any—are tripped (AL–x ON).

Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to

date. They are described in Table 3-5.