Operation 3000ZA-3X


Teledyne

Analytical

Instruments

34

V1.0

-----------------------
Entering the MODEL screen causes the above screen to be

displayed for 3 seconds. There are no modifiable parameters.

4.3.6 Self-Test Routine

-----------------------

Begin

Self-Test?

ENTER-YES

ESCAPE-NO

-----------------------
Pressing ENTER causes the analyzer to run the same diagnostic

that runs on start-up. See section 5.4 for results display.

4.3.7 Filter Routine

-----------------------

Digital

Filter

->Index:

5

-----------------------
Pressing ENTER causes the number after 'Index:' to start

flashing. Then pressing UP or DOWN selects a number ranging from 1
to 10. Lower filter values cause the analyzer to converge on new inputs
more quickly, but also allow more noise in the sample.


4.3.8 Analog-Out Adjust Routine

-----------------------

4-20mA Offset: 0