Operation 3000ZA-3X


Teledyne

Analytical

Instruments

40

any offset of the zero gas used. Pressing UP or DOWN adjusts the
number.

4.5.3 Zero Begin

-----------------------

->Zero

Begin

-----------------------
-----------------------

Begin

Zero?

ENTER-YES

ESCAPE-NO

-----------------------

Pressing ENTER causes the screen to prompt to begin the zero

routine (as displayed above). Pressing ENTER again causes the
analyzer to commence the zero routine, and begin analyzing in ZERO
MODE.

4.5.4 Zero Timer

-----------------------

->Zero Timer: 7 min.

-----------------------
This function sets the hold time in zero mode before the analyzer

actually performs the zero adjustment. Give enough time to purge the
analyzer with zero gas.