Teledyne 3000TA - General purpose trace oxygen analyzer User Manual

Page 69

background image

A-7

Trace Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

CALIBRATION GAS:

3000 series analyzer requirements for units with Auto-Cal options. The

customer must supply a control valves (or restrictors) for any SPAN or ZERO gas
source which is attached to the Auto-Cal ports. The valve should be adjusted to the
same flow rate as the sample gas . When restrictors are used, the gas pressure must
be adjusted to achieve the proper flow rate.

OPERATION WITHOUT A RESTRICTOR DEVICE:

Operation without a restrictor device is not recommend as mentioned

above. A 3000TA without any flow restrictor device was tested on 11-19-97.
This results in a flow rate of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the
standard sample sys.

REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on flow rates
at various pressures.

TAI PART NUMBERS

RESTRICTOR KIT:

A68729

UNION (SS)

U11

LP. RESTRICTOR

R2323 ( LOW PRESSURE / VAC. SERVICE )

STD.. RESTRICTOR

R2324

BLUE DOT

NUT

N73

FERRULE

F73

FERRULE

F74

BOTH FERRULES ARE
REQUIRED

CONVERSIONS:

1 PSI

=

2.04 INCHES OF MERCURY (in. Hg.)

1 SCFH

=

0.476 SLPM