background image

Trace Oxygen Analyzer

Operation 4

4-15

Teledyne Analytical Instruments

Use the

ΔΔΔΔΔ∇ arrow keys to enter the oxygen-concentration mode. Use

the < > arrow keys to blink the digit you are going to modify. Use the

ΔΔΔΔΔ∇

arrow keys again to change the value of the selected digit. When you have
finished typing in the concentration of the span gas you are using
(209000.00 if you are using air), press

Enter

to begin the Span calibration.

#### ppm Span
Slope=#### ppm/s

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates informa-
tion on Slope. Spanning automatically ends when the span output corre-
sponds, within tolerance, to the value of the span gas concentration. Then the
instrument automatically returns to the analyze mode.

4.4.2.2

Manual Mode Spanning

Press

Span

to start the

Span

function. The screen that appears allows

you to select whether the span calibration is to be performed automatically or
manually.

Span: Settling:MAN
For Next

Use the

Δ∇ keys to toggle between AUTO and MAN span settling.

Stop when MAN appears, blinking, on the display. Press

Enter

to move to

the next screen.

Press

Enter

to move to the next screen.

Calib. Holding time
Cal hold: 5 min

This menue allows the operator to set the time the analyzer should be

held in the auto span mode. It does not affect anything in Manual Mode.
Just press Enter to continue.

Span Val: 000008.00
Span Mod #

Press

Δ () to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas

you are using (209000.00 if you are using air). The < > arrows choose the
digit, and the

Δ∇ arrows choose the value of the digit.

Press

Enter

to enter the span value into the system and begin the span

calibration.