Teledyne 3000TA - General purpose trace oxygen analyzer User Manual
Page 10
1-4
1 Introduction
Model 3000TA
Teledyne Analytical Instruments
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD display:
•
Left & Right Arrows
Select between functions currently
displayed on the VFD screen.
•
Up & Down Arrows
Increment or decrement values of
functions currently displayed.
•
Enter
Moves VFD display on to the next screen in a series. If
none remains, returns to the
Analyze
screen.
•
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the
Analyze
screen.
Digital Meter Display: The meter display is a LED device that
produces large, bright, 7-segment numbers that are legible in any lighting. It
produces a continuous readout from 0-10,000 ppm and then switches to a
continuous percent readout from 1-25%. It is accurate across all analysis
ranges without the discontinuity inherent in analog range switching.
Alphanumeric Interface Screen: The VFD screen is an easy-to-use
interface from operator to analyzer. It displays values, options, and messages
that give the operator immediate feedback.
Flowmeter: Monitors the flow of gas past the sensor. Readout is 0.2 to
2.4 standard liters per minute (SLPM).
Standby Button: The Standby turns off the display and outputs,
but circuitry is still operating.
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When chassis
is exposed or when access door is open and power
cable is connected, use extra care to avoid contact
with live electrical circuits .
Access Door: For access to the Micro-Fuel Cell, the front panel swings
open when the latch in the upper right corner of the panel is pressed all the
way in with a narrow gauge tool. Accessing the main circuit board requires
unfastening rear panel screws and sliding the unit out of the case.
1.6 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.