beautypg.com

Teledyne 3000TA - General purpose trace oxygen analyzer User Manual

Page 10

background image

1-4

1 Introduction

Model 3000TA

Teledyne Analytical Instruments

Data Entry Keys: Six touch-sensitive membrane switches are used to

input data to the instrument via the alphanumeric VFD display:

Left & Right Arrows

Select between functions currently
displayed on the VFD screen.

Up & Down Arrows

Increment or decrement values of
functions currently displayed.

Enter

Moves VFD display on to the next screen in a series. If
none remains, returns to the

Analyze

screen.

Escape Moves VFD display back to the previous screen in a

series. If none remains, returns to the

Analyze

screen.

Digital Meter Display: The meter display is a LED device that

produces large, bright, 7-segment numbers that are legible in any lighting. It
produces a continuous readout from 0-10,000 ppm and then switches to a
continuous percent readout from 1-25%. It is accurate across all analysis
ranges without the discontinuity inherent in analog range switching.

Alphanumeric Interface Screen: The VFD screen is an easy-to-use

interface from operator to analyzer. It displays values, options, and messages
that give the operator immediate feedback.

Flowmeter: Monitors the flow of gas past the sensor. Readout is 0.2 to

2.4 standard liters per minute (SLPM).

Standby Button: The Standby turns off the display and outputs,

but circuitry is still operating.

CAUTION: The power cable must be unplugged to fully

disconnect power from the instrument. When chassis
is exposed or when access door is open and power
cable is connected, use extra care to avoid contact
with live electrical circuits .

Access Door: For access to the Micro-Fuel Cell, the front panel swings

open when the latch in the upper right corner of the panel is pressed all the
way in with a narrow gauge tool. Accessing the main circuit board requires
unfastening rear panel screws and sliding the unit out of the case.

1.6 Recognizing Difference Between LCD & VFD

LCD has GREEN background with BLACK characters. VFD has

DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED
.