Trace Oxygen Analyzer

Maintenance 5

5-3

Teledyne Analytical Instruments

1. Remove power to the instrument by unplugging the power cord

at the power source.

2. Open the front panel door by pressing the release button on the

top right corner of the door all the way in with a narrow gauge
tool, such as a small screwdriver, and releasing it.

3. With one hand placed underneath the cell block ready to catch

the Micro-Fuel cell , lift up on the stainless steel gate in front of
the cell block. This releases the cell and cell holder from the
block. The cell and holder will fall out in your hand.

Figure 5-1: Removing the Micro-Fuel

Cell Block

Micro-Fuel Cell

O-Ring

Cell Holder

Lift Up

Gate

Cell Adaptor

(For B-2 or A-2

series cell)