A-1

Trace Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

Appendix

A-1 Specifications

Packaging: General Purpose

• Flush panel mount (Standard).

• Relay rack mount. Contains either one or

two instruments in one 19" relay rack
mountable plate (Optional).

Sensor: Teledyne L-2 trace analysis Micro-Fuel Cell.

Cell Block: 316 stainless steel.

Sample System: All wetted parts of 316 stailess steel.

90 % Response Time: 65 seconds at 25 °C (77 °F).

Ranges: Three user definable ranges from 0–10 ppm to

0–250,000 ppm, plus air calibration range of 0-
250,000 ppm (25 %).

Autoranging with range ID output.

Alarms: One system-failure alarm contact to detect

power failure or sensor-zero failure.

Two adjustable concentration threshold alarm
contacts with fully programmable setpoints.

Displays: 2-line by 20-character, VFD screen, and one 5

digit LED display.

Digital Interface: Full duplex RS-232 communications port.

Power: Universal power supply 85-250 V ac, at

47-63 Hz.

Operating Temperature: 0-50 °C (32-122 °F)