Note:

The ranges must be increasing from low to high, for example, if range
1 is set as 0–100 ppm and range 2 is set as 0–1,000 ppm, range 3
cannot be set as 0– 500 ppm since it is lower than range 2.

Ranges, alarms, and spans are always set in ppm units (over the entire

0-250,000 ppm range), even though all concentration-data outputs change
from ppm units to percent when the concentration is above 10,000 ppm.

4.6.2 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the

analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range

function by pressing the

Range

button on the front panel.

Use the < > arrow keys to move the blinking over AUTO.

Use the

Δ∇ arrow keys to switch from AUTO to FX/LO, FX/MED, or

FX/HI to set the instrument on the desired fixed range (low, medium, or
high).

L—100 M—1000
H—10000 Mode—FX/LO

or

L—100 M—1000
H—10000 Mode—FX/MED

or

L—100 M—1000
H—10000 Mode—FX/HI

Press

Escape

to re-enter the

Analyze

mode using the fixed range.

NOTE: When performing analysis on a fixed range, if the oxygen concentra-

tion rises above the upper limit (or default value) as established by the
operator for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output continue to
read the true value of the oxygen concentration regardless of the
analog output range.

4.7

The Analyze Function

Normally, all of the functions automatically switch back to the

Analyze

function when they have completed their assigned operations. Pressing the

Escape

button in many cases also switches the analyzer back to the

Ana-