Trace Oxygen Analyzer

Installation 3

3-9

Teledyne Analytical Instruments

Table 3-2: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

3.3.3 Remote Probe Connector

The 3000TA is a single-chassis instrument, which has no Remote

Probe Unit. Instead, the Remote Probe connector is used as another method
for controlling external sample/zero/span gas valves. See Figure 3-6.

Figure 3-6: Remote Probe Connector Pinouts

The voltage from these outputs is nominally 0 V for the OFF and

15 V dc for the ON conditions. The maximum combined current that can be
pulled from these output lines is 100 mA. (If two lines are ON at the same
time, each must be limited to 50 mA, etc.) If more current and/or a different
voltage is required, use a relay, power amplifier, or other matching circuitry
to provide the actual driving current.

In addition, each individual line has a series FET with a nominal ON

resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-7.