Teledyne 306WA - Analog trace oxygen analyzer User Manual
Page 42
A-4
Appendix
Appendix
Appendix
Appendix
Appendix
Model 306W
Model 306W
Model 306W
Model 306W
Model 306WA
A
A
A
A
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
Calibration Data
Calibration Data
Calibration Data
Calibration Data
Calibration Data
Range
Range
Range
Range
Range
The ranges of this analyzer are:
Range Switch Position No. 1
PPM O
2
Range Switch Position No. 2
PPM O
2
Range Switch Position No. 3
PPM O
2
Output Signal
Output Signal
Output Signal
Output Signal
Output Signal
The output signal is D.C.
Background Gas
Background Gas
Background Gas
Background Gas
Background Gas
This analyzer is intended to measure oxygen in a background of:
The flowmeter has been set to indicate a flow of 150 cc/min. of this gas.
If any other type of gas is to be analyzed, the flowmeter must be reset for that
gas, using a displacement type flowmeter, when the flow is set to 150 cc/
min.
Calibration Table
Calibration Table
Calibration Table
Calibration Table
Calibration Table
Your calibrator will generate 0-100 ppm of oxygen.
To set calibrator: Turn calibrator switch to ON position; turn calibrator dial
to read amount of oxygen to be added.
Calibration Current (mA)
To add:
10 ppm set calibrator dial to read 100
.4
20 ppm set calibrator dial to read 200
.8
30 ppm set calibrator dial to read 300
1.2
40 ppm set calibrator dial to read 400
1.6
50 ppm set calibrator dial to read 500
2.0
60 ppm set calibrator dial to read 600
2.4
70 ppm set calibrator dial to read 700
2.8
80 ppm set calibrator dial to read 800
3.2
90 ppm set calibrator dial to read 900
3.6
100 ppm set calibrator dial to read 1000
4.0
Cell Class
Cell Class
Cell Class
Cell Class
Cell Class:
Electrolyte Type
Electrolyte Type
Electrolyte Type
Electrolyte Type
Electrolyte Type
Type A: 10% potassium hydroxide in distilled water
Type B: 10% potassium carbonate in distilled water
Type C: 20% potassium bicarbonate in distilled water