beautypg.com

Teledyne 306WA - Analog trace oxygen analyzer User Manual

Page 30

background image

TELEDYNE ELECTRONIC TECHNOLOGIES

Analytical Instruments

5-6

5.0 Maintenance &

5.0 Maintenance &

5.0 Maintenance &

5.0 Maintenance &

5.0 Maintenance & T

T

T

T

Trrrrroub

oub

oub

oub

oubleshooting

leshooting

leshooting

leshooting

leshooting

Model 306W

Model 306W

Model 306W

Model 306W

Model 306WA

A

A

A

A

adequate sensitivity, the entire cell assembly must be replaced. Refer to
section 4.2.3

4.2.3

4.2.3

4.2.3

4.2.3 for cell installation.

5.7 Calibrator

5.7 Calibrator

5.7 Calibrator

5.7 Calibrator

5.7 Calibrator

The calibrator must be clean in order to operate properly. It should be

removed and inspected each time the cell is serviced. If, on inspection,
dark deposits are seen on the calibrator wick, the calibrator should be
replaced. See section 4.3: Calibrator

4.3: Calibrator

4.3: Calibrator

4.3: Calibrator

4.3: Calibrator.

The scrubber is not repairable. Once it has been determined that it is

not functioning correctly, it should be replaced. To determine if the scrub-
ber is active or not:

1. Allow sample gas with a few PPM of O

2

into the sensor. Record

the output registered on the meter

2. Turn the scrubber on and allow zero gas into the sensor. Record

the output.

3. If the output in 1 and 2 are the same, the scrubber needs to be

replaced.

5.8 Reservoir and Humidifier Column

5.8 Reservoir and Humidifier Column

5.8 Reservoir and Humidifier Column

5.8 Reservoir and Humidifier Column

5.8 Reservoir and Humidifier Column

Approximately once each year the reservoir and humidifier column

should be drained and cleaned. Use the following procedure:

a) Reduce sample flow to approximately 50 cc/min

50 cc/min

50 cc/min

50 cc/min

50 cc/min.

b) Refer to Figure 6. Place a small funnel with attached tubing

beneath the drain spout which is located on the underside of the
reservoir. Remove the drain spout cap and allow the reservoir
and humidifier column to drain thoroughly.

c) Replace the drain spout cap, turn the sample flow off, and

remove the fill cap on the top of the reservoir.

d) Add a small amount of electrolyte to a pint of warm distilled

water. Pour the solution into the reservoir and replace the fill
cap.

e) Gently open the throttle valve and permit the sample to flow for

five to ten minutes.

f) Reduce the sample flow to 50 cc/min

50 cc/min

50 cc/min

50 cc/min

50 cc/min. and drain the solution

from the reservoir. When drained, replace the drain cap and
turn off the sample flow.

g) Rinse by filling with distilled water and draining several times.