Teledyne 306WA - Analog trace oxygen analyzer User Manual

Page 38

background image

TELEDYNE ELECTRONIC TECHNOLOGIES

Analytical Instruments

5-14

5.0 Maintenance &

5.0 Maintenance &

5.0 Maintenance &

5.0 Maintenance &

5.0 Maintenance & T

T

T

T

Trrrrroub

oub

oub

oub

oubleshooting

leshooting

leshooting

leshooting

leshooting

Model 306W

Model 306W

Model 306W

Model 306W

Model 306WA

A

A

A

A

Symptom

Symptom

Symptom

Symptom

Symptom

Cause

Cause

Cause

Cause

Cause

What

What

What

What

What T

T

T

T

To Do

o Do

o Do

o Do

o Do

Calibrator current is erratic.

Proper calibration current exists,
but there is no cell response.

a) Poor electrical connections.

a) Short circuit in calibrator.

a) Check for proper

connections between
the control unit and
calibrator on terminal
strip TS5.

b) Ensure that brown (13)

and yellow (14) wires
connect to brown and
yellow posts respec-
tively on calibrator
assembly in the cell
compartment.

c)

Moisten wick. See
section 4.3.

a) Verify cell response to

oxygen by loosening the
calibrator assembly to
introduce air, and then
bolt back in place.
Recorder/meter should
show increase.

b) If there is no response,

inspect the calibrator for
obvious shorts. If not
easily repairable,
replace the calibrator.