Teledyne 306WA - Analog trace oxygen analyzer User Manual
Page 28
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
5-4
5.0 Maintenance &
5.0 Maintenance &
5.0 Maintenance &
5.0 Maintenance &
5.0 Maintenance & T
T
T
T
Trrrrroub
oub
oub
oub
oubleshooting
leshooting
leshooting
leshooting
leshooting
Model 306W
Model 306W
Model 306W
Model 306W
Model 306WA
A
A
A
A
2.
2.
2.
2.
2.
Quick Disconnect Fitting Method
Quick Disconnect Fitting Method
Quick Disconnect Fitting Method
Quick Disconnect Fitting Method
Quick Disconnect Fitting Method
NOTE:
NOTE:
NOTE:
NOTE:
NOTE: Only stainless steel tubing must be connected to the quick discon-
Only stainless steel tubing must be connected to the quick discon-
Only stainless steel tubing must be connected to the quick discon-
Only stainless steel tubing must be connected to the quick discon-
Only stainless steel tubing must be connected to the quick discon-
nect fitting.
nect fitting.
nect fitting.
nect fitting.
nect fitting.
The steps are as follows:
- Connect 1/4" tubing with the male quick disconnect fitting to the
Model 306 sample inlet.
- Turn on the sample gas.
- Adjust the flow control inside the analyzer until the ball is inside the
target (on the flow tube).
- Purge the analyzer for 12 to 16 hours when the instrument is being
installed for the first time or when replacing the cell.
- Remove the quick disconnect from sample inlet and connect it to the
span inlet.
- Slowly adjust the span gas regulator until span gas flow is about the
same as the sample flow.
- Wait until the reading is stabilized. Adjust span pot so that the
analyzer is reading the same O
2
concentration as indicated on the
span gas bottle.
- Now the analyzer is calibrated.
-Remove the quick disconnect from span inlet and connect it to the
sample inlet and the instrument is ready to go online.
5.5 Cell
5.5 Cell
5.5 Cell
5.5 Cell
5.5 Cell
The electrochemical reaction in the analytical process results in the
accumulation of lead ions in the electrolyte, making the replacement of the
lead electrode or the electrolyte necessary.
5.5.1 Electrolyte Replacement
5.5.1 Electrolyte Replacement
5.5.1 Electrolyte Replacement
5.5.1 Electrolyte Replacement
5.5.1 Electrolyte Replacement
Teledyne recommends that the electrolyte be replaced when foreign
material is accumulated in the cell or when the cell sensitivity is noticeable
decreased. This is typically identified when a large change is made to the
span pot setting during calibration. Remove the cell from is compartment,
and drain, clean, rinse, and refill as described in section 4.2.3
4.2.3
4.2.3
4.2.3
4.2.3. After the
cell is serviced or replaced, calibrate the analyzer as outlined in Section
4.7.1: Calibration Using the Internal Calibrator
4.7.1: Calibration Using the Internal Calibrator
4.7.1: Calibration Using the Internal Calibrator
4.7.1: Calibration Using the Internal Calibrator
4.7.1: Calibration Using the Internal Calibrator.