Turbine Generator Gas Analyzer

Maintenance

Teledyne

Analytical

Instruments

37

analyzer) in the calibration
mode.

check and recalibrate if
necessary.

or

(b) Improper zero source. Check

source.

or

(c) Improper range selected—

select proper range for zero.

No sample flow

(a) Sample flow valve closed—

open the valve.

or

(b) Pressure at regulator set too

low—reset the regulator to 0.5-
30 psig

or

(c) No sample gas connected—

connect sample gas.

or

(d) Vent or sample in line is

blocked. Remove constriction
or blockage.

Analyzer does not respond to
sample gas

(a) Incorrect instrument range

selected. Select proper range
for analysis.

or

(b) Make sure CAL switch is in

Analyze mode.

or

(c) No sample flow—check

flowmeter.