Teledyne 2750 - Portable turbine generator gas analzyer User Manual

Page 36

background image

Maintenance Model

2750

Teledyne

Analytical

Instruments

36

If a leak detector is available, the model can be leak checked with

the power off. Use helium and pressurize the lines to 1-5 psig and sniff
all fittings and lines.

If a leak detector is not available, admit air at 1-5 psig through the

instrument lines. Liberally apply a commercial soap bubble solution to
all fittings and lines. Most leaks occur at the gas inlet fitting.

CAUTION:

DO NOT USE THIS INSTRUMENT UNTIL IT
SUCCESSFULLY PASSES A LEAK CHECK TEST.

5.2 Troubleshooting

Table 5-1 describes common problems and corrective action to be

taken.

Table 4-1: Troubleshooting

Symptom Corrective

Action

Unit does not turn ON when
actuating Power

(a) Check DC adapter is mated in

securely to the port connector.

or

(b) Check DC Adapter is plugged

into 90-264VAC 47-63 Hz
power.

or

(c) Power adapter is defective—

replace adapter.

Heater lamp does not illuminate
or illuminates continuously

(a) No power to the unit—see above.

or

(b) Unit cannot attain proper

temperature due to wind or
temperature too cold—move
unit out of wind/draft or to a
warmer area.

Display reads negative (with
zero gas flowing through

(a) ZERO control may be

misadjusted. Perform a zero