Turbine Generator Gas Analyzer

Setup and Operation

Teledyne

Analytical

Instruments

31

Setup and Operation

4.1 Setup

Setup consists of:

 Calibrating the instrument
 Attaching the sample and vent lines
 Leak checking
 Attaching power cable
 Attaching output cable
 Turning the instrument on and allowing it to warm up

Prior to using the Model 2750, it must be calibrated on all three

ranges. Refer to Section 3.3 for calibration procedures.

Once the instrument is calibrated, it may be transported to the

analysis site.

Note: It is sometimes more convenient to calibrate the instrument

at the analysis site.

Open the cover and attach the sample line to the sample valve using

the 1/8” tubing connector. Attach a suitable vent line heeding the
precautions stated in Section 3.2.1. Leak check thoroughly all
connections and tubing using a commercially available soap solution for
this task.

CAUTION:

HYDROGEN IS A FLAMMABLE AND EXPLOSIVE
GAS. NEVER LEAVE ANYTHING TO CHANCE WHEN
USING HYDROGEN. ALWAYS LEAK CHECK THE
SYSTEM PRIOR TO USE.

Plug the 9V DC adapter into the power port and the other end to a

90-264 VAC power outlet. Plug in the 0-1 VDC output cable into the
output port if you intend to record the data from the analyzer. See
Figure 4-1.