Setup and operation – Teledyne 2750 - Portable turbine generator gas analzyer User Manual
Page 31

Turbine Generator Gas Analyzer
Setup and Operation
Teledyne
Analytical
Instruments
31
Setup and Operation
4.1 Setup
Setup consists of:
Calibrating the instrument
Attaching the sample and vent lines
Leak checking
Attaching power cable
Attaching output cable
Turning the instrument on and allowing it to warm up
Prior to using the Model 2750, it must be calibrated on all three
ranges. Refer to Section 3.3 for calibration procedures.
Once the instrument is calibrated, it may be transported to the
analysis site.
Note: It is sometimes more convenient to calibrate the instrument
at the analysis site.
Open the cover and attach the sample line to the sample valve using
the 1/8” tubing connector. Attach a suitable vent line heeding the
precautions stated in Section 3.2.1. Leak check thoroughly all
connections and tubing using a commercially available soap solution for
this task.
CAUTION:
HYDROGEN IS A FLAMMABLE AND EXPLOSIVE
GAS. NEVER LEAVE ANYTHING TO CHANCE WHEN
USING HYDROGEN. ALWAYS LEAK CHECK THE
SYSTEM PRIOR TO USE.
Plug the 9V DC adapter into the power port and the other end to a
90-264 VAC power outlet. Plug in the 0-1 VDC output cable into the
output port if you intend to record the data from the analyzer. See
Figure 4-1.