beautypg.com

Teledyne 2750 - Portable turbine generator gas analzyer User Manual

Page 23

background image

Turbine Generator Gas Analyzer

Installation

Teledyne

Analytical

Instruments

23

3.3.1 Zero Calibration Procedure

1. Connect the zero gas (100% CO

2

) to the sample valve. Make

sure that the pressure regulator is set at approximately 5 psig.

CAUTION:

ALWAYS HAVE THE SAMPLE FLOW VALVE
CLOSED WHEN TURNING ON THE SAMPLE (OR
CALIBRATION GAS) STREAM. FLOW RATES
GREATER THAN 2.5 SCFH CAN DAMAGE THE
SENSOR.

2. If not already connected, connect the vent line.

3. Plug the DC adapter into the unit. See Figure 3-2. Plug the other

end into a 90-264 VAC 43-67 Hz supply.

4. Turn the instrument on and allow it to warm up for 20 minutes.

5. Allow the zero gas to flow through the analyzer. Wait several

minutes before making any flow adjustments. Then set the gas
flow to 0.5-1 SCFH using the flow control valve.

6. With CO

2

flowing through the instrument, set the calibration

switch (see Figure 3-3) to the air/CO

2

position for zeroing that

range. Verify that the range selector switch is on the air/CO

2

position.

Figure 3-3: Interface Panel Showing Calibration Switch Position