Turbine Generator Gas Analyzer

Installation

Teledyne

Analytical

Instruments

21

there are no kinks or obstructions in the return line to avoid over
pressurizing and possibly damaging the analyzer.

Figure 3-2: Gas Connections

3.2.1 Vent Line

Gases exiting the instrument should be vented to an area remote

from the analyzer. This line must not be blocked! The following
precautions must be observed in vent line installation:

1. Use 3/16” ID tubing to eliminate any backpressure or restricted

flow.

2. Vent to an area where the ambient pressure is normal

atmospheric pressure.

3. The ambient pressure in the vent area must undergo no more

than normal barometric pressure changes.

4. The vent line must be installed in such a manner as to avoid

water and dirt accumulation.

3.2.2 Pressure Regulation

All incoming gas lines from a high-pressure source should be

equipped with pressure regulators. Pressure regulators should be
installed as close to the sample take off point as possible to minimize