beautypg.com

Installation – Teledyne 2750 - Portable turbine generator gas analzyer User Manual

Page 19

background image

Turbine Generator Gas Analyzer

Installation

Teledyne

Analytical

Instruments

19

Installation

The Model 2750 is a portable unit and can easily be setup close to

the analysis site. To ensure accuracy, however, the analyzer should be
setup in such a manner as to avoid:

 Direct sunlight
 Drafts of air
 Shock and vibration
 Temperatures below 30° F or above 110° F

For best results, the analyzer should be placed as close as possible

to the sample point to minimize the effects of sample line lag time on the
analysis.

An outline diagram, showing the location and identification of the

gas port and electrical connections, as well as the physical dimensions of
the analyzer case, is included in the drawings at the rear of the manual.

Installation is quick and easy and consists of:

 Power and signal output connection
 Connecting the sample or calibration gas
 Allowing the analyzer to warm up.
 Calibrating the instrument.

3.1 Electrical Connections

Power to the instrument is supplied by a 90-264 VAC 47-63 Hz

universal external power adapter. This adapter provides 1.4 Amperes at
9 VDC required by the instrument. The adapter plugs into a port on the
left side of the main panel as shown in Figure 3-1.

A second port, also on the left side of the panel, provides access to

a 0-1 VDC (0.8 – 1 VDC for the H

2

in Air range) output signal for

percent of range monitoring or recording. A connector is supplied for
interfacing to this port. To use this signal, connect a 2-conductor