Theory of Operation

Model 2750

Teledyne

Analytical

Instruments

18

width modulated (PWM) signal that is proportional to the gas
concentration on that specific range.

The PWM signal is passed through a low pass filter and is used to

drive the display and the 0-1 VDC output signal (0.8 – 1VDC for the
H

2

/Air range).

The calibration switch is a rotary switch and allows either the

percent of range signal or the output from the zero offset gain amplifiers
to be sent to the display.