Turbine Generator Gas Analyzer

Setup and Operation

Teledyne

Analytical

Instruments

33

2. Adjust the pressure regulator to 0.5-30 psig.

3. Slowly open the sample valve and adjust the flow on the

flowmeter to 0.1-1 SCFH.

Note: The operator must know in advance what gas mixture is

present in the sample stream to be able to properly select
the instrument range. All ranges will give some indication
of output (even if zero) but only that range corresponding
to the correct sample gas mixture will give an accurate
reading.

4. Make sure the calibration switch is in the analyze position

5. Select the proper range and allow gas to flow for several minutes

to purge the lines of any residual air or gas.

Note: Whenever switching gases, always allow a sufficient purge

time to remove the residual gas. Note the changing
concentration values until a steady reading is obtained
(unless the sample gas stream concentration is changing).
When the reading settles, the instrument is sufficiently
purged.

6. Read the sample gas concentration from the display.